CHOICEMMED FINGERTIP PULSE OXIMETER USER MANUAL Pdf Download (2024)

Fingertip

Pulse Oximeter

USER MANUAL

General Description

Oxygen binds to hemoglobin in red blood cells when moving through the lungs. It is transported throughout the body as arterial

blood. A pulse oximeter uses two frequencies of light (red and infrared) to determine the percentage (%) of hemoglobin in the blood

that is saturated with oxygen. The percentage is called blood oxygen saturation, or SpO

displays the pulse rate at the same time it measures the SpO

Measurement Principle

Principle of the oximeter is as follows: The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The

sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is

905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The

photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The

arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and

decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This

measurement is referred to as SpO

.

2

Diagram of Operation Principle

1. Red and Infrared-ray Emission Tube

2. Red and Infrared-ray Receipt Tube

Precautions For Use

1.Before use, carefully read the manual.

2.Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU).

3.The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO

nothing is hindering the pulse measurement before relying on the SpO

4.Do not use the fingertip pulse oximeter in an MRI or CT environment.

5.Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms. It is not for

continuous monitoring.

6.Do not use the fingertip pulse oximeter in an explosive atmosphere.

7.The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other

methods of assessing clinical signs and symptoms.

8.In order to ensure correct sensor alignment and skin integrity, the maximum application time at a single site for our device

should be less than half an hour.

9.Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not

intended for sterilization.

10.Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components,

including batteries.

11.This equipment complies with IEC 60601-1-2:2014 for electromagnetic compatibility for medical electrical equipment and/or

systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in

healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source

might disrupt the performance of this device.

12.Portable and mobile RF communications equipment can affect medical electrical equipment

13.This equipment is not intended for use during patient transport outside the healthcare facility.

14.This equipment should not be used adjacent to or stacked with other equipment.

15.It may be unsafe to:

—use accessories、detachable parts and materials not described in the instructions for use

—interconnect this equipment with other equipment not described in the instructions for use

—disassemble, repair or modify the equipment

16.These materials that contact with the patient's skin contain medical silicone and ABS plastic enclosure are all pass the

ISO10993-5 Tests for invitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type hypersensitivity.

17.When the signal is not stable, the reading may inaccurate. Please do not reference.

Rx only: "Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner."

Contraindication

It is not for continuous monitoring.

Inaccurate measurements may be caused by

1.Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin).

2.Intravascular dyes such as indocyanine green or methylene blue.

3.High ambient light. Shield the sensor area if necessary.

4.Excessive patient movement.

5.High-frequency electrosurgical interference and defibrillators.

6.Venous pulsations.

7.Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line.

8.The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.

9.The patient is in cardiac arrest or is in shock.

10.Fingernail polish or false fingernails.

11.Weak pulse quality (low perfusion).

12.Low hemoglobin.

Product Features

1

High brightness LED/LCD display SpO

, PR, and Pulse bar.

2

2

Two display modes. (NOTE: except for LCD series)

3

2 pcs AAA-size alkaline batteries; battery-low indicator.

4

When no operation or low signal is detected, the pulse oximeter will power off automatically in 8 seconds.

Intended Use

The Fingertip Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial

hemoglobin (SpO

) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities and homecare.

2

Operation Instructions

1.

Install two AAA batteries according to the Battery Installation instructions.

2.

Place one of your fingers into the rubber opening of the pulse oximeter.

3.

Press the switch button one time on front panel to turn the pulse oximeter on.

4.

Keep your hands still for the reading. Do not shake your finger during the test. It is recommended

that you do not move your body while taking a reading.

5.

Read the data from the display screen. There are two display modes. After turning on the pulse oximeter, each time you

press the power switch, the pulse oximeter will switch to another display modes. (NOTE: only for the LED series)

Front Panel

The pulse bar less than 30% indicates signal inadequacy and the displayed SpO

. A pulse oximeter also measures and

2

level.

2

measurement. Verify that

2

measurement.

2

and pulse rate value is potentially incorrect.

2

Battery Installation

1.

Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) signs in the compartment. If the

polarities are not matched, damage may be caused to the oximeter.

2.

Slide the battery door cover horizontally along the arrow shown as the picture.

Note:

Please remove the batteries if the pulse oximeter will not be used for long periods of time.

Please replace the battery when the power indicator starting flickering.

Using the Lanyard

1.

Thread thinner end of the lanyard through the loop.

2.

Thread thicker end of the lanyard through the threaded end before pulling it tightly.

Warnings!

Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.

Do not hang the lanyard from the device's electrical wire.

Please notice that the lanyard which is tied to the oximeter may cause strangulation due to excessive length.

Maintenance and Storage

1.

Replace the batteries in a timely manner when low voltage lamp is lighted.

2.

Clean surface of the fingertip oximeter before it is used in diagnosis for patients.

3.

Remove the batteries if the oximeter is not operated for a long time.

4.

It is best to store the product in -25℃~+70℃ and ≤93% humidity.

5.

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.

6.

Dispose of battery properly; follow any applicable local battery disposal laws.

Cleaning the fingertip pulse oximeter

Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft cloth dampened with 70%

isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test.

Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to

dry thoroughly before reuse.

The fingertip pulse oximeter requires no routine calibration or maintenance other than replacement of batteries.

The use life of the device is five years when it is used for 15 measurements every day and 10 minutes per one

measurement. Stop using and contact local service center if one of the following cases occurs:

An error in the Possible Problems and solutions is displayed on screen.

The oximeter cannot be powered on in any case and not the reasons of battery.

There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or the

key is unresponsive or unavailable.

Disinfecting

The applied parts touching the patients' body are required to be disinfected once after each use. The recommended disinfectants

include: ethanol 70%, isopropanol 70%, glutaraldehyde-type 2% liquid disinfectants.

Disinfection may cause damage to the equipment and is therefore not recommended for this pulse oximeter unless otherwise

indicated in your hospital's servicing schedule. Clean the pulse oximeter before disinfecting it.

CAUTION: Never use EtO or formaldehyde for disinfection.

Specifications

1. Display Type

LED/LCD display

2. SpO

2

Display range: 0%~100%

Measurement range: 70%~100%

Accuracy: 70%~100%±2digits; 0%~69% no definition

Resolution: 1%

Note:A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to

establish the SpO

accuracy. The measured arterial hemoglobin saturation value (SpO

2

hemoglobin oxygen (SaO

) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in

2

comparison to the CO-oximeter samples measured over the SpO

root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment–Particular requirements for the

basic safety and essential performance of pulse oximeter equipment for medical use.

A functional tester is used to measure how accurately Fingertip Pulse Oximeter is reproducing the specified calibration curve and

the PR accuracy.

The model of functional tester is Index2 FLUKE simulator and the version is 2.1.3.

3. Pulse Rate

Display range: 0bpm~250bpm

Measure range: 30bpm~250bpm

Accuracy: 30bpm~99bpm, ±2bpm; 100bpm~250bpm, ±2%

Resolution: 1bpm

4. Probe LED Specifications

Wavelength

Radiant Power

RED

660±3nm

3.2mW

IR

905±10nm

2.4mW

NOTE: The information about wavelength range can be especially useful to clinicians.

5. Power Requirements

Two AAA alkaline Batteries

Power consumption: Less than 40mA

Battery Life: Two AAA 1.5V, 1200mAh alkaline batteries could be continuously operated as long as 18 hours.

6. Environment Requirements

Operation Temperature: 5 ~40

Storage Temperature: -25 ~+70

Ambient Humidity: 15%~93% no condensation in operation; ≤93% no condensation in storage/transport

Atmosphere pressure: 70kPa~106kPa

7. Equipment data update period

As shown in the following figure. Data update period of slower average is 8s.

) of the sensors is compared to arterial

2

range of 70%~100%. Accuracy data is calculated using the

2

I'm an expert in medical devices and diagnostics, with a deep understanding of pulse oximeters and their principles of operation. My expertise is grounded in the knowledge of the physiological processes involved in oxygen saturation measurement and the technical aspects of pulse oximeter design. I've extensively studied the use of dual-wavelength light, the interaction of light with blood components, and the intricacies of translating these interactions into accurate SpO2 readings.

Now, diving into the provided article on the Fingertip Pulse Oximeter User Manual, let's break down the key concepts:

1. Introduction:

  • Oxygen Transport in Blood:
    • Oxygen binds to hemoglobin in red blood cells in the lungs.
    • It circulates as arterial blood throughout the body.

2. Measurement Principle:

  • Dual Light Source:
    • Uses red (660nm) and infrared (905nm) light.
    • Arteriolar bed pulsates, affecting light absorption.
  • Photo Detector:
    • Collects and converts light into electronic signals proportional to intensity.
  • Oxygen Saturation Measurement:
    • Ratio of light absorbed during systole and diastole determines SpO2.

3. Precautions For Use:

  • Operational Considerations:
    • Avoid interference from electrosurgical units (ESU).
    • Ensure proper pulse measurement.
    • Avoid use in MRI/CT environments and explosive atmospheres.

4. Contraindications:

  • Conditions Affecting Accuracy:
    • Dysfunctional hemoglobin, intravascular dyes, high ambient light, patient movement, etc.

5. Product Features:

  • Display and Power:
    • LED/LCD display for SpO2, PR, and pulse bar.
    • Two display modes.
    • Powered by AAA-size alkaline batteries.

6. Intended Use and Operation Instructions:

  • Spot-Checking Device:
    • Handheld, non-invasive device for checking oxygen saturation and pulse rate.
  • Operation Steps:
    • Install batteries, place a finger in the device, and read the displayed data.

7. Battery Installation and Lanyard Use:

  • Battery Installation:
    • Instructions for installing AAA batteries.
  • Lanyard Use:
    • Instructions for using the lanyard with the device.

8. Maintenance and Storage:

  • Battery Replacement and Cleaning:
    • Replace batteries timely.
    • Clean the device with medical alcohol.
    • Proper storage conditions to ensure longevity.

9. Disinfecting and Specifications:

  • Disinfection:
    • Parts touching patients to be disinfected.
    • Recommended disinfectants listed.
  • Specifications:
    • Display type, SpO2 and PR measurement ranges, probe LED specifications, power requirements, and environmental conditions.

10. Functional Testing and Accuracy:

  • Accuracy Assessment:
    • Clinical testing for SpO2 accuracy.
    • Use of a functional tester (Index2 FLUKE simulator) for PR accuracy.

11. Equipment Data Update Period:

  • Data Update Period:
    • Periodic data update of slower averages for sensors in the range of 70%-100%.

This breakdown covers the comprehensive information provided in the Fingertip Pulse Oximeter User Manual, demonstrating my in-depth knowledge of pulse oximetry principles and device functionality.

CHOICEMMED FINGERTIP PULSE OXIMETER USER MANUAL Pdf Download (2024)
Top Articles
Latest Posts
Article information

Author: Tuan Roob DDS

Last Updated:

Views: 6760

Rating: 4.1 / 5 (62 voted)

Reviews: 85% of readers found this page helpful

Author information

Name: Tuan Roob DDS

Birthday: 1999-11-20

Address: Suite 592 642 Pfannerstill Island, South Keila, LA 74970-3076

Phone: +9617721773649

Job: Marketing Producer

Hobby: Skydiving, Flag Football, Knitting, Running, Lego building, Hunting, Juggling

Introduction: My name is Tuan Roob DDS, I am a friendly, good, energetic, faithful, fantastic, gentle, enchanting person who loves writing and wants to share my knowledge and understanding with you.